Antithrombin III assay
Summary
The turbidity of the immunocomplex produced by Antithrombin-III (AT-III) and its corresponding specific antibody is measured by transmission method, and its turbidity is directly proportional to the concentration of AT-III in the serum. This experiment is derived from the experimental instruction of the undergraduate 5-year professional laboratory of Mudanjiang Medical College.
Operation method
Antithrombin III Content Measurement Test
Principle
The turbidity of the immunocomplex produced by Antithrombin-III (AT-III) and its corresponding specific antibody was measured by a transmission method, and its turbidity was proportional to the concentration of AT-III in the serum.
Move
1. Manual, semi-automatic determination
Data processing: AT-Ⅲ content (mg / L) - (measured tube A value / reference tube A value) X reference concentration.
Normal value: 210-360 dL
Caveat
1. When the concentration of the specimen is too high, the dilution ratio of the specimen should be adjusted and re-measured.
2. high blood fat. Jaundice and mild hemolysis have little effect on the results of this method.
3. The linear range of this method is 0-560mg/L-.
4. The parameters can be adjusted according to the different models of automatic analyzers.Clinical significance: High AT-III content: seen in hemophilia, aplastic anemia, heart valve disease, uremia, renal transplantation, etc. Decreased AT-III content: seen in DIC, liver disease, toxemia thrombosis, oral contraceptives and other diseases. Oral contraceptives and other diseases.